Dual Mode - °F & °C
(For Clinical Use)

High Precision

Easy to Read LCD Display

Fast Response

Easy to use

Alarm Signal

Measurement in °F & °C

Auto shut-off

Break Resistance

HOW TO USE

  1. Press Power switch to turn on the unit.
  2. Place the thermometer under the tongue or the underarm & wait for beeper sound.
  3. Read Temperature
  4. Press power switch to turn off the unit.
  5. Before taking Oral, Axillary or Rectal temperature, please read the enclosed pack insert.

Products Description


Specification

Range : 32.0°C - 42.9°C (90.0°F) (°C/°F chosen by manufacturer)
Accuracy ±0.1°C, 35.5°C - 42.0°C (±0.2°F, 95.9°F-107.6°F)
±0.2°C, under 35.5°C or over 42.0°C
±0.4°F, under 95.9°F or over 107.6°F
at standard room temperature of 25°C (77.0°F)
Display Liquid crystal display, 3½ digits
Memory for storing the last measured value
Battery One 1.5 V DC. button battery (size SR41, UCC 392 or LR41)
Power consumption 0.15 milliwatts in measurement mode
Battery Life Approx 200 hours
dimension 12.4cm x 1.8cm x 0.9cm (L x W x H)
Weight Approx. 10 grams including battery
Using environment Temperature: 10°C~35°C (50°F~95°F)
Relative humidity: 25% RH~80% RH
Storage and transportation condition Temperature: 25°C~55°C (-13°F~131°F)
Relative humidity: 25% RH~80% RH

How to use:

  1. Wipe or dip probe with ethyl alcohol for disinfection.
  2. Push the on/off button. The LCD will show for 2 seconds and you will hear a beep simultaneously. Last tested temperature will be shown for about 2 seconds.  Then LCD shows the self-test temperature of 37.0°C (98.6°F). After that Lo°C (Lo°F) sign shows.
  3. Position the probe in mouth, rectum, armpit etc.
  4. Once the temperature-talking is finished, the degree sign of °C (°F) on the will stop flashing, and you will hear buzzes for approx.10 seconds. The reading will not charge. The tested temperature is shown on LCD. During or after the test, if the test temperature is <32.0°C (or 90.0°F), the LCD shows Lo °C (or Lo °F); If the tested temperature is extra high, the LCD shows Hi °C (or HI °F). Consult doctor at once. After temperature-taking, if the tested temperature is <37.8°C (100.0°F), the buzzes are “Bi- Bi- Bi- Bi-”; if the tested temperature is ≥37.8°C (100.0°F), the buzzes whose frequency are much higher are “Bi-Bi-Bi---- Bi-Bi-Bi---- Bi-Bi-Bi”.
  5. The thermometer will automatically turn off in about minutes for energy-saving. However, to prolong battery life, it is suggested to press the On/Off button to turn the thermometer off once test is finished.
  6. Please dispose exhausted battery and thermometer according to the relevant law.

Note: In step 2, if the room temperature is higher than 32°C (90°F), LCD will display the room temperature, instead of Lo °C (or Lo °F).

Note: use of the probe cover may result in a 0.1°C (0.2°F) discrepancy from actual temperature.

ORAL USE

Put the probe into either position marked by “ √ ” as showed in the picture right-hand for about 1 minute. To insure the accuracy, please keep the mouth close for 5minutes before taking temperature. The normal temperature by this method is 35.7°C-37.3°C (96.3°F-99.2°F).

RECTAL USE

Lubricate the probe with the petroleum jelly for easier insertion. Gently insert the about one centimeter (Less than ½ inch.) Into the patient’s rectum and keep for about 1 minute. The normal temperature by this method is 36.2°C-37.7°C (96.3°F – 99.9°F).

AXILLARY USE

From a medical viewpoint, this method will always provide inaccurate readings. Wipe armpit with a dry towel. Place probe into armpit and keep the arm firmly against the body side for about 1.5 minutes. The normal temperature by this method is 35.2°C - 36.7°C (95.4°F – 98.1°F).

CLEANING AND DISINFECTION

  1. Clean thermometer by wiping it with a dry cloth and sterilize the probe with ethyl alcohol.
  2. Don’t let thermometer contact any chemical thinner.
  3. Not completely waterproof. Do not soak the LCD in water. Only the probe is waterproof.

PRECAUTION

Do not drop thermometer. It is not shock proof.
Do not bend thermometer.
Do not boil the probe.
Do not store thermometer in direct sunlight.
Do not attempt to disassemble thermometer except replacing the battery.
  Means direct volts D.C.
  Means type B.

Battery Replacement

  1. When the “ ” appears in the lower right corner of the LCD, the battery is exhausted and needs replacement.
  2. Pull the battery cover off with hand in the direction shown below.
  3. Gently pull out the circuit board with battery chamber approximately 1 cm.
  4. Use pointed object such as a pen to take out the exhausted battery. Discard exhausted battery out reach of children.
  5. Place a new 1.5V D.C. button size battery type SR41, UCC 392 or LR41 or equivalent in the chamber with positive side faced up and negative side faced down.
  6. Cover the battery.

LIMITED WARRANTY

The thermometer is guaranteed for 90 days from the date of purchase. If thermometer does not function properly due to defective component or poor workmanship, we will repair or replace it free of charge. All components are covered by this warranty excluding the battery.